ABSTRACT
Objective:
Cricoid pressure is useful in fasted patients requiring emergency intubation. We compared the effect of cricoid pressure on laryngeal view during Macintosh, McGrath MAC X-Blade and GlideScope video laryngoscopy.
Methods:
After obtaining approval from the Human Research Ethics Committee and written informed consent from patients, we enrolled 120 patients (American Society of Anesthesiologists I-II, age 18-65 years) undergoing elective surgery that required endotracheal intubation in this prospective randomised study. Patients were divided into three groups (Macintosh, McGrath MAC X-Blade and GlideScope).
Results:
Demographic and airway variables were similar in the groups. Cormack-Lehane grades were improved or unchanged on using cricoid pressure in Macintosh and McGrath MAC X-Blade groups. However, laryngeal views worsened in 12 patients (30%), remained unchanged in 26 patients (65%) and improved in 2 patients (5%) in the GlideScope group (p<0.001). Insertion and intubation times for Macintosh and McGrath MAC X-Blade video laryngoscopes were similar. Insertion times for GlideScope and Macintosh video laryngoscopes were similar, but were longer than those for the McGrath MAC X-Blade video laryngoscope (p=0.02). Tracheal intubation took longer with the GlideScope video laryngoscope than with the other devices (p<0.001 and p=0.003). Mean arterial pressures after insertion increased significantly in Macintosh and GlideScope groups (p=0.004 and p=0.001, respectively) compared with post-induction values. Heart rates increased after insertion in all three groups compared with post-induction values (p<0.001). Need for optimisation manoeuvres and postoperative minor complications were comparable in all three groups.
Conclusion:
Although all three devices are useful for normal or difficult intubation, cricoid pressure improved Cormack-Lehane grades of Macintosh and McGrath MAC X-Blade video laryngoscopes but statistically significantly worsened that of the GlideScope video laryngoscope.