Gnana Laryngeal Airway in Clinical Practice: A Prospective Observational Study

Turk J Anaesthesiol Reanim 2020;48(4):280-287
1. Department of Anaesthesia and Intensive Care, Government Medical College and Hospital, Chandigarh, India
2. Valley Anaesthesiology and Pain Consultants, Phoenix, AZ; Creighton University School of Medicine - Phoenix Regional Campus, Phoenix, AZ; University of Arizona College of Medicine-Phoenix, Department of Anesthesiology, Phoenix, AZ, USA
3. Department of Anaesthesiology and Pharmacology, Toxicology, and Neurosciences, Louisiana State University School of Medicine, Shreveport, Louisiana, USA
4. Beth Israel Deaconess Medical Center, Department of Anaesthesia, Critical Care, and Pain Medicine, Harvard Medical School, Boston, MA, USA
No information available.
No information available
Received Date: 02.06.2019
Accepted Date: 15.08.2019
Publish Date: 26.12.2019
PDF
Cite
Share
Request

ABSTRACT

Objective:

The Gnana Laryngeal Airway (GLA) device, a novel supraglottic airway device, is similar to the LMA-Classic in basic design, but with an additional suction port on the convex portion of the laryngeal mask to remove the saliva. We evaluated the GLA device in terms of ease and time to insertion, the number of attempts, oropharyngeal leak pressure (OLP), correct placement, and complications in adult patients undergoing elective surgical procedures.

Methods:

After general anaesthesia, the GLA device was inserted in ASA Class I–II consecutive patients aged 18–60 years, who were scheduled for elective surgeries lasting <2h. An independent observer noted (1) 10 consecutive successful GLA device insertions, all on the first attempt; (2) 10 consecutive device insertions, each <20 second in duration; and (3) 10 consecutive patients with the mean leak <10%. The criteria were fulfilled in 50 consecutive patients.

Results:

In 72% of patients, the GLA device was successfully placed on the first attempt and was effortless in 64%. Between the first 10 and last 10 consecutive patients of the total 50, the ease-of-insertion grade progressively decreased (mean±standard deviation [SD]: 2.80±0.25 to 1.30±0.15, p<0.0001) and so did insertion time in seconds (28.70±1.87 to 14.20±0.79, p<0.0001). The post-insertion, OLP and airway compliance progressively increased, while the cuff inflation volume, peak airway pressure and airway resistance progressively decreased, along with minimal side effects and malposition.

Conclusion:

The GLA device insertion became progressively easier and faster; thus, such a device is promising and warrants further clinical evaluation.

Keywords: Compliance, ease of insertion, Gnana Laryngeal Airway, oropharyngeal leak pressure, peak airway pressure, resistance